Test kit AC-6 for acidity determination

SKU: 222301
Manufacturer: Hach
Availability: Order only
The AC-6 set includes:
- volumetric sample tube
- mixing bottle
- reagents (see Accessories tab)
- instructions
- carrying case
Description Specifications Recommended products Accessories
Description
Description
Acidity test recommendations
The presence of chlorine in the water sample may slightly interfere with the test. Up to 50 mg/L chlorine can be removed with 0.1 N. sodium thiosulfate solution, cat. No. 323-37. This solution is not included in the kit. Add one drop of 0.1N sodium thiosulfate solution to the sample before adding the indicator. Over-agitation of the sample should be avoided to prevent the loss of dissolved gases such as carbon dioxide, hydrogen sulfide, or ammonia.
Free Acidity Test - High Range (0 - 400 mg/L CaCO3)
- Fill the volumetric tube with sample water. Pour the contents into a mixing bottle.
- Add 1 sachet of reagent (94399) to the mixing container and shake to mix.
- If the water turns gray-blue, blue, or green, the free acidity is zero.
- If the water turns pink, add Sodium Hydroxide Standard Solution (2349332) dropwise while holding the pipette vertically. Count each drop as you add. Mix the sample after each drop. Continue adding drops until the water turns gray-blue.
- The free acidity of the water, expressed in mg/L CaCO3, is equal to the number of drops of sodium hydroxide standard solution added multiplied by 20.
Acidity determination - Low range (0 - 100 mg/L CaCO3)
- Fill the mixing bottle with sample up to the 23 ml mark.
- Add 1 sachet of reagent (94399) to the mixing container and shake to mix.
- If the water turns gray-blue, blue, or green, the acidity is zero.
- If the water turns pink, add Sodium Hydroxide Standard Solution (2349332) dropwise while holding the pipette vertically. Count each drop as you add. Stir the mixing bottle after each drop. Continue adding drops until the water turns gray-blue.
- The acidity of water in mg/L, as calcium carbonate (CaCO3), is determined by multiplying the added number of drops of a standard sodium hydroxide solution by 5.
Step-by-step determination of total acidity (0 - 400 mg/L CaCO3)
- Fill a volumetric tube with a sample of water. Pour the contents into a mixing container.
- Add one drop of Phenolphthalein Indicator Solution (189736) to the mixing bottle and shake to mix.
- If the water turns pink, the total acidity is zero.
- If the water remains colorless, add sodium hydroxide standard solution (2349332) dropwise, holding the pipette vertically. Count each drop as you add. Stir the mixing bottle after each drop. Continue adding drops until the sample is slightly pink.
- The acidity of the water, expressed in mg/L as CaCO3, is equal to the number of drops of sodium hydroxide standard solution used, multiplied by 20.
Specifications
| Application area | Environmental monitoring, Wastewater, Treatment facilities, Laboratory analysis, Process control |
| Industry | Municipal |
| Parameters | Acidity |
| Measurement Range | 5 - 100 mg/L CaCO3 |
| Measurement method | drop count titration |
| Method | Sodium hydroxide |
| Number of tests | 100 |
Recommended products
Digital titrator reagent kit, chloride 10 - 8000 mg/L Cl
Titrimetric determination of chlorides in waste water by the method of mercury nitrate (method Hach 8206).
SKU: 2272600
SKU: 1601-0046
SKU: XM Octopus™
SKU: 1340-5110
SKU: DULCODOS Ammonia
SKU: 1340-6621
SKU: DULCOFLEX DFCa
SKU: 1601-0047
SKU: DUODOS
SKU: 1202009S
SKU: 1340-5415
SKU: 1340-5739
Digital titrator reagent kit, carbon dioxide 10 - 1000 mg/L CO2
For the determination of carbon dioxide by titration on a digital titrator. Hach Lange method 8205. The kit is sufficient for 40 -100 tests.
SKU: 2272700
Test kit, total chlorine, model CN-DT, 20 - 2000 mg/L
Test kit, total chlorine; range 20 - 2000 mg/L Cl2.
SKU: 2471100
R4, R8, R12 - Laboratory Mash Bath
The mash bath is designed for use in laboratories of breweries and malt houses. It is designed to determine the extract content of malt using the CONGRESS, HARTONG (20, 45, 65, 80) and ASBC methods. Users can also carry out tests using their own algorithms using the programmable programs TERMOSTAT or PROFILE.
SKU: R4, R8, R12
SKU: CPDI0NJM1
SKU: DULCOSMOSE NF
SKU: Bello Zon CDVd
Test set SU-5 for the determination of sulfite
The kit contains 100 tests that use drop titration / iodometric analysis to determine sulfite.
SKU: 148002
SKU: 1340-6201
SKU: 1601-0020
Hach titration kit: pH and total acidity (milk, tomato sauce, fruit juice)
Hach accessory kit (without reagents) for pH and total acidity determination in the food industry.
SKU: AP0008.AT1102
SKU: 1340-2700
Potentiometric titrator Titralab AT1222, 2 Burettes, 2 Pumps
Automatic potentiometric Hach titrator for fast and accurate titrations.
SKU: AT1222.82
Accessories
Bromcresol green - methyl red indicator, powder pillows, 100 pieces
Powder Pillows for titration (alkalinity).
SKU: 94399
Phenolphthalein indicator solution; 1 g/L, 15 ml
Solution for use in titration.
SKU: 189736
Standard solution, sodium hydroxide, 0.035 N, 100 ml
Solution for determining the acidity of a water sample.
SKU: 2349332
Recommended products
Titralab AT1000 Series Potentiometric Titrator, 1 Burette
Titralab AT1000 Series Potentiometric Titrator for fast and accurate titration.
SKU: AT1102.98
SKU: MAKRO M5Ha
SKU: Bello Zon CDVd
SKU: DULCOSMOSE BW
SKU: 1340-5415
Distillation unit UDK 129
The UDK 129 is a basic but powerful system for a variety of tests, including the measurement of various types of nitrogen (e.g. ammonia and protein Kjeldahl or direct alkaline distillation), as well as the analysis of substances such as phenols, volatile fatty acids, cyanide and alcohol.
SKU: UDK 129
SKU: 1202009S
SKU: OZONFILT OZVb
Hach Titration Kit: Ca / Mg Water Hardness (ISE)
Set of accessories (without reagents) Hach for determination of calcium and magnesium hardness of water.
SKU: AP0003.AT1112